¤ýAnimec AM-2S
 
General Description

The AM-2S is an electrically powered, dry-heat warmer which supplies external heat to the plastic tubing incorporated in intravenous administration sets. The IV Tube set is placed in the shaped channel in the aluminum heating plate. The device is available in 4mmØ (diameter 3.1 ~ 4.0) and 5mmØ (diameter 4.1 ~ 5.0) models for all tubing sizes.

Operating temperature
27¢ªC-37¢ªC(80.6¢ªF- 98.6¢ªF) *
Flow rate:
1~12ml / min.
Safety Features
Heating plate temperature sensor, fluid temperature sensor, heater sensor, temperature fuse, splash resistant construc-tion, dry-heat
Size/weight 176 (L) x 65 (W) x 36 (H) mm; 560g (1.2lbs)
* Effluent temperature is affected by initial fluid temperature, room temperature, flow rate and diameter of the tube set used.

Name and function of each component
 
 
¤ýBenfits
 
  1) Automatic regulation
The temperature of the warming plate adjusts according to the desired temperature of the effluent.
  2) No disposables
Special warming bags or coils are not required.
 
  3) Less heat loss after warming
Small size neccessitates minimal tubing between the patient and the unit preventing cooling of the infused fluid.
 
  4) Quick set-up
The tube is inserted directly into the warmer; no priming or disposable preparation is required.
     
  5) Rapid warm-up
Ready to use in 2 ~ 3 minutes
 
  6) Safe performance
Protection from overheating, electric shock and contamination.
     
  7) High durability
Simple construction and shock resistant casing.
   
 
Flow rate vs fluid temperature
 
 
Fiud temperature changes due to influence of room temperature
 
 
Initial Fluid Temperature
15¡ÆC ¡¾ 1¡ÆC
Flow Rate [ml / min]
3 5 8 10 12 15

Distance from the exit of unit

100mm 35.5 34.2 30.8 29.4 28.1 26.5
250mm 31.5 32.2 30.1 29.0 27.7 26.3
400mm 29.1 30.7 29.2 28.0 27.1 26.2
600mm 26.7 28.3 28.9 28.0 27.1 26.1
 
 